Pipetting and operators: a study of 108 people
A 2026 study weighed the water pipetted at 10 and 100 µL by 108 people, with pipettes recalibrated within the previous 15 days. The results, the authors' recommendations and the difference between calibration and technique.
The calibration of a pipette tells how the instrument delivers, under controlled conditions. In everyday work the result also depends on the person using it. A study published in 2026 in The Journal of Applied Laboratory Medicine measured how much, in 108 people who work or train in laboratories.
The study
The study by Ö. Özcan, P. J. M. Grundel, A. C. Heijboer and W. P. J. den Elzen involved 108 participants: 26 laboratory school students, 10 PhD students, 10 clinical chemists and 62 technicians, who were further divided into three groups according to how much sensitive manual pipetting their work involves.
26
Laboratory school students
10
PhD students
10
Clinical chemists
62
Technicians, in three groups
Each participant dispensed ten consecutive aliquots of 10 µL and ten of 100 µL of purified water, weighed on a balance, and performed a serial dilution of a copper sulfate solution. The tests were carried out at 23.5 °C and 52 % relative humidity, with water left at ambient temperature for 24 hours. The same pipettes were used for all experiments, recalibrated within the previous 15 days. Results were compared with the maximum permissible errors of ISO 8655:2022.
The results
Accuracy outside the ISO limits
10 %
of participants at 10 µL
22 %
of participants at 100 µL
Imprecision outside the ISO limits
36 %
of participants at 10 µL
20 %
of participants at 100 µL
- Median coefficient of variation between 0.3 and 0.8 % at 100 µL and between 1.4 and 5.9 % at 10 µL, according to the summary published by ADLM.
Students showed the highest variability and the longest times; technicians in the group with the most sensitive tasks achieved the best dilution. Experienced staff, in some cases, did not stay within the limits of the standard either. The authors conclude that experience and task sensitivity significantly affect performance, and recommend ongoing training and periodic assessment for all laboratory staff, with particular attention to low volumes.
Calibrating the instrument, assessing the technique
The study separates two sources of error that add up in a laboratory's results. The pipettes, the same for everyone, had been recalibrated within the previous 15 days: the differences between participants come from technique, not from the instrument.
The instrument
Calibration concerns the instrument. ISO 8655-6 specifies the gravimetric method as the reference procedure: the instrument delivers water into a vessel on a balance, the delivered mass is weighed and converted into volume, and water and room temperature, relative humidity and atmospheric pressure are recorded. On adjustable-volume instruments at least three volumes are tested, with at least ten measurements per volume; on multichannel pipettes each channel is tested as a single-channel instrument. For each volume the certificate gives the systematic error and the random error, with their uncertainty.
The person
Technique concerns the person, and it is assessed with the same means used in the study: water, a balance, repeated readings. Calibration does not correct the operator's technique.
- Ö. Özcan, P. J. M. Grundel, A. C. Heijboer, W. P. J. den Elzen, "Pipetting Performance across Laboratory Personnel: A Comparative Study", The Journal of Applied Laboratory Medicine 11(3):460–469, 2026.
- ADLM, Clinical Laboratory News, "Experience matters in pipetting", 6 May 2026.
- Accredited calibration of balances and micro-dispensers: methods and tests, test volumes, reading the certificate.
- The LAT calibration certificate for micro-dispensers: certificate contents and parts of ISO 8655.
- Pipette and micro-dispenser calibration: what changes with the new ISO 8655-6:2022.